Abstract
The guidewire (GW) is probably the most important tool for the endovascular treatment of arterial obstructions. In fact, the treatment of a lesion is only possible when the GW is passed beyond the target lesion. Lower limb percutaneous arterial revascularisation can be achieved using a variety of GWs which may differ in calibre, body, tip or coating. As the choice of an appropriate GW is critical for the success of a lower limb artery angioplasty, knowledge of the properties and performances of different GWs should be well known in order to tailor the choice of the device to the lesion characteristics and location, as well as to the intended revascularisation strategy. The aim of the present paper is to describe the constructive characteristics of GWs for lower limb arterial revascularisation, and to evaluate groups of GWs for each segment of the lower limb arterial vasculature.
Introduction
Lower limb arterial endovascular revascularisation is increasingly performed worldwide due to the rising prevalence of peripheral artery disease as well as the improvement in equipment and techniques. Guidewires (GWs) are necessary tools for any type of vascular angioplasty, allowing lesion crossing and advancement of devices such as balloons and stents beyond a lesion1. Until recently, the choice of GWs specifically designed for lower extremity artery angioplasty has been limited. For example, there were no dedicated below-the-knee (BTK) GWs and such procedures were frequently performed using coronary GWs. Currently, the spectrum of GWs designed for lower limb interventions has broadened but their characteristics have not been adequately summarised. The aim of this paper is to describe the constructive properties and performance of GWs dedicated to lower limb arterial interventions in order to help operators to choose the best device according to lesion characteristics and the segment affected.
Properties of guidewires
The functional properties of a GW depend on its calibre and on the constructive characteristics of the tip and the body. The characteristics of the tip are the main determinants for crossing the lesion, while the characteristics of the body are critical for the delivery of devices (e.g., balloons and stents) at the level of the lesion.
WIRE CALIBRE
GWs are first classified according to their calibre. The most widely used GWs in the lower limb arterial vasculature are 0.014”, 0.018” or 0.035” in diameter.
As a general rule, the larger the calibre, the greater the support, torque control and pushability; and, the smaller the calibre, the greater the flexibility and trackability of the GW. The term “crossing GW” can be applied to a GW specifically designed to traverse challenging lesions. The principal features of a crossing GW are a high penetration capacity and a good torque control of the tip. Often these GWs have a hydrophilic tip.
After having crossed the lesion with a crossing GW, per definition quite “aggressive”, it is advisable to exchange it for a “position GW” whose principal features are a soft, non-hydrophilic tip coupled with a supportive body, to allow equipment advancement while minimising the risk of damaging the distal vasculature.
The choice of calibre of the GW must be made according to the access site and the type as well as the localisation of the lesion treated: this affects the size of all the devices used in the procedure. In fact, balloons and stents must be compatible with the GW (and the sheath) used. As a general rule, 0.035” devices are compatible with a 6 Fr or larger sheath, 0.018” devices can be used with a 4 or 5 Fr sheath, and 0.014” devices are for the most part compatible with a 4 Fr sheath strategy. Larger calibre devices on a smaller GW (e.g., a 0.035” compatible device advanced over a 0.018” GW or a 0.018” compatible device advanced over a 0.014” GW) can be used but navigation may be more difficult.
WIRE TIP
The tip characteristics of a 0.035” GW are limited to their general shape (i.e., straight, curved), the length of the flexible part and the coating (i.e., hydrophilic or not). The tip characteristics of 0.014” and 0.018” GWs include the design, the load and the coating.
GWs can have a shaping ribbon design, i.e., the core does not reach the tip of the wire, or a core-to-tip design, i.e., the core reaches the tip (Figure 1A, Figure 1B). The core-to-tip design permits a better torque control but renders the wire more aggressive (i.e., with a greater risk of vessel perforation). Usually, the diameter of a GW is the same throughout its length. However, some GWs have a progressive reduction of the diameter near the tip. This characteristic increases the penetration capacity. These devices are called “tapered tip” GWs (Figure 1C).
The tip load of a GW is calculated with a tester that measures the force that has to be applied to the last 10 mm of the tip to flex it 2 mm laterally. This measure is expressed in grams and gives an indication of the penetration capacity of a GW. However, as the penetration power is inversely correlated to the area of the tip of the wire, tapered GWs have a far higher penetration capacity than an equal tip load, non-tapered device.
The cover of the tip also influences the ability of the GW to cross a lesion. In particular, polymeric coating gives a hydrophilic property to the tip that becomes slippery once in contact with the bloodstream. These hydrophilic GWs are more suitable to cross tight and calcific stenosis, although they carry a higher risk of perforation. The curve of the tip of a GW has to fit the diameter of the vessel to be treated and the bends of the arterial tree in which the device has to navigate. Therefore, a proper shaping of the tip is important to enhance the performance of a GW.
Lower limb artery diameters span from 10-12 mm of a common iliac artery to 1.5-2 mm of a below-the-ankle vessel. While 0.035” hydrophilic peripheral GWs, used mainly in large vessels, are frequently pre-shaped, most 0.018” and 0.014” peripheral GWs have a straight-tip, shapeable design.
The accuracy with which the tip of a GW can be shaped depends mainly on its inner composition. A steel core of the tip allows an accurate shaping of the tip, but subsequently the degree of the curve can only be increased. Therefore, the tip of a steel core GW has to be shaped with care and step by step, because it is nearly impossible to reverse the curves. GWs with a nitinol core are more difficult to shape because they tend to recover their original profile; however, once modified, they can be re-shaped. The tip of the GW can be shaped using the inserter or with the non-cutting edge of the blade (Figure 2).
The final bend of the tip of the GW can be a gentle, continuous curve or can comprise single or multiple acute bends (Figure 3). Acute bends allow a more precise tip control but may be difficult to obtain in GWs with a nitinol core. Therefore, curvilinear bends can be adequate for 0.035” GWs with a nitinol core and appropriate for aortoiliac and femoropopliteal interventions. On the other hand, acute bends are more easily created in 0.018” and 0.014” GWs with a steel core and are more appropriate for BTK interventions.
WIRE BODY
A steel core gives good torque control, pushability and support, but it is more prone to be irreversibly damaged. These characteristics are ideal for a position GW. A nitinol core is associated with less torque control, pushability and support, but the wire is more kink resistant. According to the different characteristics of steel and nitinol, the core of the body of some GWs is now made of steel in the most proximal part, to have more support and torque control, and of nitinol in the more distal part, to offer flexibility and durability (e.g., Command; Abbott Vascular, Santa Clara, CA, USA). An alternative solution for a similar result is obtained in hybrid nitinol GWs: the proximal part is a super stiff shaft due to a non-hydrophilic spiral coating for better torque control and device delivery, while the most distal part is a standard hydrophilic tip for better crossing capacity (e.g., Glidewire Advantage®; Terumo Corp., Tokyo, Japan).
The distance from the access site to the target lesion and the bends the GW has to pass through during its path both influence the choice of wire with respect to its body structure. Accordingly, for procedures via the arm or via a transfemoral crossover access, the support of the wire is critical and a steel core may be preferred. For an antegrade BTK angioplasty, the durability of a nitinol core may be of advantage.
In addition, the body of the GW can be covered with a hydrophilic coating that becomes slippery when wet. This constructive property confers trackability to the device.
The transition point between the body and the tip is also important. In fact, when attempting a subintimal tracking, the wire will flex at the transition point, which, therefore, should not be too far from the tip to allow a thinner loop.
Principles of lesion crossing
In lower extremity artery obstructive disease, lesions are heterogeneous in terms of vessel size, type of lesion (i.e., stenosis or occlusions), length of lesion, and degree of calcification. It is critical to differentiate GWs dedicated to lesion crossing from those allowing balloon and stent advancement once the lesion is crossed. The choice of GW calibre must be guided only by the crossing issue. Regardless of the vessel size (aortoiliac, femoropopliteal, BTK vessels) and in every vascular area, any size can be used (i.e., 0.014”, 0.018” or 0.035” calibre), with the sole aim of getting into the true distal lumen (Figure 4). GWs must be tailored to the different situations that can be encountered.
In stenotic lesions, a residual true lumen with antegrade blood flow is present. Therefore, it is of paramount importance to maintain the endoluminal position during lesion crossing, avoiding subintimal tracking, especially in the presence of vessel tortuosity. In these cases, a “sliding” strategy must be pursued: a soft-tip GW may be rotated and slowly advanced in order to avoid producing any dissection. A flexible tip GW is preferred, while a hydrophilic coating can help in trackability (e.g., Radifocus® Guidewire M Standard [Terumo], AQUATRACK™ Regular [Cordis, Johnson & Johnson, Warren, NJ, USA] as 0.035” GWs [Table 1]; Connect [Abbott Vascular], V-18™ Long Taper [Boston Scientific, Marlborough, MA, USA] as 0.018” GWs [Table 2]; Command, V-14™ Long Taper [Boston Scientific] as 0.014” GWs [Table 3]). Configuring the tip shape of smaller calibre GWs to progress within the vessel pathway is also important, and an acute, 1-2 mm long bend is usually adequate for 0.018” and 0.014” GWs.
In crossing occlusions the endoluminal approach is in any case the first choice irrespective of the length of the lesion2. The first GW to be used can be a hydrophilic 0.014” or 0.018” GW with a flexible tip, trying to slide into the occluded arterial lumen following the softer pathways inside the occluding material. In case of failure of this approach, a “perforating strategy” using a more aggressive GW specifically designed to traverse hostile (long and/or calcified) lesions is recommended. The principal features of these GWs are the high penetration capacity of the tip and good torque control. Such wires are often non-hydrophilic and have tapered tips with a high tip load (e.g., Connect 250T [Abbott Vascular], Astato 30 [ASAHI Intecc, Aichi, Japan], Victory™ 18 30 g [Boston Scientific] as 0.018” GWs; Winn 200T [Abbott Vascular], Astato XS 20 [ASAHI Intecc], Victory™ 14 30 g [Boston Scientific] as 0.014” GWs [Table 1-Table 3]). Notably, the ability of a wire to cross the lesion is enhanced by advancing a balloon or a support catheter over the wire near its tip.
As the endoluminal approach has a high rate of failure in long occlusions, the subintimal approach represents an alternative option for the endovascular treatment of these lesions (Figure 5)3. The characteristics of an ideal GW for a subintimal crossing are a hydrophilic tip and high body support, with the flexing point located close to the tip. In fact, the short flexing tract permits the creation of a small loop and the support of the wire assures its pushability, while the hydrophilic distal tract acts as a cutting blade. The 0.035” GW most used for subintimal recanalisation is the Radifocus® M Stiff (Terumo), with the flex point at 3 cm from the tip (Table 1). A GW with similar characteristics is the AQUATRACK™ Stiff (Cordis) (Table 1). When attempting a subintimal approach in BTK vessels, it is better to switch to a 0.018” or 0.014” GW. The 0.018” GWs most used for subintimal tracking are the V-18™ (Boston Scientific), short taper version, where the flex point is 8 cm from the tip, and the Connect Flex (Abbott Vascular) with the flexing point 3 cm from the tip (Table 2). In case of subintimal tracking using a 0.014” GW, the Command ES (Abbott Vascular) can be used (Table 3).
In subintimal recanalisation the most challenging moment is the re-entry into the true lumen. Although specifically designed (and expensive) re-entry devices do exist, the first attempt to re-enter into the true lumen is usually made with the same GW used to progress within the subintimal space. However, when the GW used to dissect fails to find its downstream way, a second attempt with a different GW can be made. The technical characteristics of a “re-entry” GW can be summarised as follows: first, the smaller the better. In fact, it is often advisable to downsize the re-entry GW with respect to the calibre of the dissecting GW. For example, it is worth trying to re-enter the distal SFA with a 0.018” GW after subintimal tracking and a failed re-entry with a 0.035” GW, or to re-enter a distal tibial artery with a 0.014” GW after a failed attempt with a 0.018” GW. Second, a non-hydrophilic GW can be superior to a hydrophilic one, as it is less prone to progress into the dissection. Third, a GW with good torque control is needed because the tip has to be directed precisely towards the true lumen or within a small hole communicating with the true lumen. Lastly, a GW with a high-grade tip load (tapering is also helpful) is necessary to perforate the intima and gain access to the true lumen (Figure 5).
The above-mentioned characteristics can be found in the Astato 30 or the Connect 250T (Table 2) when using 0.018” GWs, or the Astato XS 20 (Table 3) when the choice is a 0.014” GW. Before being used, the tip of a re-entry GW should be shaped adequately. A 45 degree angulation of the tip should be obtained (Figure 3). The length of the angulated portion should fit the vessel to be re-entered, i.e., the larger the target vessel, the longer the angulated tract. Therefore, the angulated portion should be 3-4 mm long when trying to re-enter within an iliac artery, 2-3 mm long when re-entering a distal SFA or 1-2 mm long when the vessel to be re-entered is a tibial artery; a secondary bending can also be added to lean the GW against the outward wall of the vessel and increase its back support (Figure 3). When in the subintimal space with the re-entry GW, a proper and accurate view should be sought with different angulations of the flat panel (also with the help of calcifications if present) in order to point the tip of the GW exactly towards the true lumen. As soon as in the true lumen, after having verified the position by angiography with a balloon catheter, these very aggressive GWs should be replaced by a position GW.
Principles of lesion treatment
After having crossed the lesion, one should reconsider whether the GW used for lesion crossing is adequate for the next steps of the procedure, namely, balloon angioplasty and possibly stenting. If not appropriate, the crossing GW should be exchanged for a position GW, according to vessel size and anatomy. When choosing a position GW it is important to take into account its calibre, tip load and the body support.
In fact, it is possible to cross an iliac lesion with a 0.014” GW and, after predilatation, exchange it for a 0.035” position GW for final treatment. On the other hand, it is possible to perform a subintimal dissection of a tibial artery with a 0.035” GW and exchange it for a 0.014” position GW for subsequent dilatations. Similarly, dedicated GWs for crossing hostile occlusions have a very stiff tip that can damage the distal vessels. The final treatment can be carried out using a less aggressive “position GW” whose principal feature is a soft, non-hydrophilic tip. Finally, if a flexible GW fails to give adequate support for the advancement of balloons and stents at the lesion level, a “position GW”, with a high-grade body support, will facilitate equipment advancement.
Therefore, the best characteristics of a GW for the treatment of a lesion are a soft non-hydrophilic tip and a very supportive body. When possible, and taking into account the size of balloons and stents (4, 5, 6 Fr compatible) in use, it is also useful to upgrade the calibre of the GW.
Guidewires for aortoiliac angioplasty
Suprainguinal angioplasty can be carried out retrogradely via ipsilateral femoral access or anterogradely via the contralateral femoral artery (crossover) or from the upper extremities, via brachial or radial arteries4. In this territory, an intraluminal approach should be pursued as much as possible because subintimal tracking carries the risk of life-threatening perforations.
When facing a stenosis in this tract, especially when working aligned with the vessel, i.e., retrogradely by ipsilateral femoral access or from the arm, either a 0.014”, a 0.018” or a 0.035” GW may be used. In fact, the aim is to navigate within the lumen beyond the stenosis. This can be accomplished using a hydrophilic GW with a high body support. Examples among 0.035” GWs include the AQUATRACK Stiff (Table 1), a hydrophilic GW with a nitinol core, and the Radifocus M Stiff, straight or curved-tip (Terumo), which has similar features (Table 1). Alternative 0.018” GWs include the V-18 Long Taper and the Connect Flex, while amongst 0.014” GWs the Command ES may be used.
When dealing with an iliac occlusion, it may be wise first to attempt to cross intraluminally to reduce the risk of perforation. First attempts can be carried out with a hydrophilic flexible-tipped GW (e.g., Radifocus M Standard [Terumo] or AQUATRACK Regular as 0.035” GWs, Table 1; Connect, V-18 Long Taper as 0.018” GWs, Table 2; Command, V-14 Long Taper as 0.014” GWs, Table 3). In case of failure, non-hydrophilic, tapered 0.014” GWs (e.g., Astato XS 20, Table 3) or 0.018” GWs (e.g., Astato 30 or Connect 250T, Table 2) are all valuable choices because they have good torque control and are less prone to progress subintimally. Hydrophilic 0.018” (e.g., V-18 Short Taper or Connect Flex, Table 2) or 0.035” (e.g., AQUATRACK Stiff or Radifocus M Stiff, Table 1) GWs may also be used; however, these slippery devices definitely tend to dissect and progress subintimally.
When intraluminal tracking appears impossible, a subintimal crossing may be attempted with the above-mentioned 0.018” or 0.035” hydrophilic GWs. For the subintimal tracking of an iliac occlusion, an approach from above (either in crossover from the contralateral femoral artery or from the arm) is safer than a retrograde ipsilateral one because of a lower risk of dissecting the aortic wall.
Hainwise contains other products and information you need, so please check it out.
Once the iliac stenosis or occlusion is crossed, the GWs may be replaced by a position GW. This is extremely important when entering from a brachial or a radial artery or when working in crossover. In fact, strong support of the GW is mandatory due to the long distance from the entry site to the target point when working from the arm, or because the devices need to cross a steep angle, when working in crossover. For this purpose, the 0.035” Supra Core® (Abbott Vascular), the Stiff Shaft (Covidien, Plymouth, MN, USA) and the Amplatz Super Stiff® (Boston Scientific) GWs are all adequate options (Table 1). Position 0.018” GWs include the Steelcore (Abbott Vascular), the Platinum Plus® (Boston Scientific) and the Plywire (Optimed, Ettlingen, Germany), the latter being particularly useful when working from the arm because it is 400 cm long (Table 2). The stiff body of these GWs permits an effective tracking of the devices, while the floppy, non-hydrophilic distal part guarantees safe positioning beyond the lesion. Upsizing the calibre of the crossing GW is possible after predilatation, using a larger calibre balloon as an exchange catheter.
Examples of guidewires used in the aortoiliac segment are shown in Figure 4.
Guidewires for femoropopliteal angioplasty
In this vascular segment, when facing a stenosis, even if subocclusive, a 0.018” GW may be the first choice, and the V-18 Short Taper (Table 2) may be a valuable option, as it possesses a hydrophilic tip with an 8 g tip load. Moreover, its high-grade support permits the progression of balloons and stents (and also of 0.035” compatible devices). A good alternative with similar characteristics is the Connect Flex (Table 2).
Intraluminal recanalisation should also be attempted in SFA occlusions to minimise the risk of perforation. An attempt with the above-mentioned 0.018” hydrophilic GWs may be tried first, while non-hydrophilic, tapered GWs such as the Astato 30 or the Connect 250T (Table 2) could be a more effective choice in case of initial failure. As mentioned, a balloon or a support catheter positioned near the tip of the GW enhances the penetration ability of the device. However, especially in the long, calcific occlusion of the SFA, it is frequently impossible to progress within the true lumen. For this reason, Amman Bolia3 in found that a subintimal new route to the distal SFA might represent an effective alternative. For subintimal angioplasty in the femoropopliteal tract, both 0.035” and 0.018” GWs may be used. Probably the 0.035” GW most used for subintimal angioplasty of the SFA is the Radifocus M Stiff, while the AQUATRACK Stiff has similar characteristics (Table 1). Among the choices of 0.018” GWs for subintimal SFA angioplasty is the V-18 Short Taper version or the Connect Flex (Table 2).
When the distal lumen is gained, position is best maintained with a position 0.035” or 0.018” GW, chosen from among the ones suggested for aortoiliac interventions.
When all antegrade attempts fail, a retrograde access from the SFA, distal to the re-entry point, may be attempted. A popliteal and, more recently, a distal femoral puncture5 have been suggested for this purpose. In this case, the GW has to navigate retrogradely, re-enter proximally upstream from the occlusion, and be captured by a snare or within a catheter into the sheath, to complete the angioplasty anterogradely. For that purpose either the V-18 Short Taper or the Connect Flex may be used (Table 2).
Examples of guidewires used in the femoropopliteal segment are shown in Figure 5.
Guidewires for below-the-knee angioplasty
The GWs most used for BTK angioplasty are 0.014” ones, although some operators prefer 0.018” GWs, and, in selected cases, even 0.035” GWs may be used2. Nonetheless, coronary GWs have been frequently utilised in the past and are still utilised nowadays for BTK angioplasty (Table 4).
The choice of GW for BTK angioplasty mainly depends upon the characteristics of the lesion (i.e., stenosis versus occlusions, presence versus absence of calcifications) and the strategy the operator decides to pursue (i.e., intraluminal versus subintimal, antegrade versus retrograde recanalisation).
In addition, in BTK interventions, an intraluminal GW advancement should be pursued as much as possible. For this purpose, a 0.014” hydrophilic GW may be the first choice. These characteristics can be found in the Command ES or in the Victory family of GWs (Boston Scientific) (Table 3). These GWs, using a “sliding” strategy with delicate rotations and pushing, may succeed in finding a “soft inner pathway within the occluded arterial lumen, surrounded by stiffer walls”2. In case of failure, if needing to pursue a “perforating” approach, a stiffer, non-hydrophilic, tapered 0.014” GW like the Astato XS 20 or the Winn 200T (Table 3) can be used to take advantage of their penetration ability. Intraluminal progression can also be pursued by upgrading to 0.018” hydrophilic GWs (e.g., V-18 Short Taper or Connect Flex, Table 2) and subsequently, according to a step-by-step approach, to 0.018” non-hydrophilic, tapered GWs (e.g., Astato 30 or Connect 250T, Table 2).
Using these aggressive GWs the risk of perforating the vessel increases. Therefore, once the lesion is crossed, they should be replaced by a position GW using a support catheter or a balloon dilatation catheter. The characteristics of a position GW are similar to those mentioned previously, i.e., it should have a soft, non-hydrophilic tip and a high body support. Valuable 0.014” position GWs may be the Spartacore (Abbott Vascular) or the Platinum Plus 0.014”; alternative 0.018” position GWs include the Steelcore and the Platinum Plus 0.018”, long taper version.
If the antegrade intraluminal approach fails, a subintimal wire tracking can be attempted, with great attention to not damaging a landing zone for a distal bypass. For subintimal tracking, the previously cited Command ES or one of the Victory 0.014” hydrophilic GWs may be used, although 0.018” GWs with a hydrophilic tip coating and a stiff body such as the V-18 Short Taper or the Connect Flex may be even more effective. Also, 0.035” hydrophilic GWs can be used for subintimal BTK angioplasty. For this purpose, the Radifocus M Half-stiff (Terumo) may be a good compromise between the effectiveness and the risk of perforation (Table 1, Table 2, Table 3). In BTK subintimal interventions, it is particularly important to maintain the loop as short and thin as possible because these vessels are small in diameter and re-entry is more difficult with larger loops.
When the antegrade progression of the GW cannot be achieved and a retrograde passage to the proximal true lumen has to be found, the pedal-plantar loop technique6, the transcollateral approach7 and the retrograde puncture are all potentially effective strategies8,9.
The pedal-plantar loop technique is based on the possibility of navigating from dorsal to plantar circulation and vice versa through the connections of the plantar arch of the foot. Therefore, when one tibial artery is occluded and the antegrade recanalisation has failed, it is possible to try a retrograde recanalisation through the other patent tibial artery and the plantar arch. As this technique means navigating within thin and tortuous vessels, and because damaging the distal circulation can be limb-threatening, the GWs to be used in this technique should have good torque control and a hydrophilic but soft tip. Moreover, as balloons have to be advanced through the plantar arch to the distal tract of the opposite tibial artery, good support of the GW is also needed. GWs with these characteristics include the 0.014” Command and the V-14 Long Taper (Table 3).
When trying a transcollateral tracking via the perforating branches of the peroneal artery to the distal tract of the anterior or posterior tibial arteries, a soft-tip GW with less support is needed. The 0.014” Regalia XS 1.0 (ASAHI) and the Command may both be adequate (Table 3).
When performing a retrograde recanalisation of the tibial arteries, either with the puncture of the dorsalis pedis or the distal posterior tibial artery, a GW with good support is necessary. In fact, it is mandatory to try the rendezvous in the proximal part of the occluded vessel with the GW alone or, at most, with a balloon or a support catheter. The GW frequently chosen for these attempts is the 0.018” V-18 Short Taper (Table 2)2.
Examples of guidewires used in the BTK segment are shown in Figure 6.
Conclusions
The choice of an appropriate GW is critical for the success of a lower limb arterial revascularisation. The properties and performance of each GW should be well known to the interventionalist in order to tailor the choice of GW to the lesion characteristics and location as well as to the revascularisation strategy.
Conflict of interest statement
R. Ferraresi is a consultant for Abbott Vascular, Cook Medical, Medtronic, Terumo Medical Corporation, Cordis, and Biotronik. M. Manzi is a scientific board member for Abbott Vascular, and a proctor for Boston Scientific, Bard Medical, Cook Medical, and Medtronic. M. Roffi has received institutional research grants from Abbott Vascular, Biotronik, Biosensors, Boston Scientific, and Medtronic. R. Lorenzoni has no conflicts of interest to declare.
Critical limb ischemia (CLI), characterized by nonhealing extremity ulcers and rest pain, represents the most severe form of peripheral artery disease (PAD). Patients diagnosed with CLI are often faced with the grim prospect of limb amputation.1 Those who undergo limb amputation often have a morbid prognosis with high 1-year mortality. CLI involves a variety of complex pathologies and is technically challenging to treat, requiring interventionists to be familiar with a variety of tools and techniques. Patients with CLI often present with multilevel disease extending both above and below the knee, chronic total occlusions (CTOs), and calcified arteries and lesions, which adds complexity to the case.2
The PRIME registry3 is the first CLI registry in the United States with the primary goal of understanding disease patterns in CLI patients. This registry recently examined the demographics of the first 328 patients and showed that 51% of the lesions were categorized as CTOs, 30.3% of the population had isolated below-the-knee lesions (eg, tibials), and 45.2% of the population had multilevel disease (ie, above- and below-the-knee lesions). To manage this complex pathology, a thorough understanding of wire characteristics provides a foundation for the endovascular interventionist to personalize care and develop an algorithm for success. A wide range of guidewires are available to interventionists. Length, diameter, gram strength, and hydrophilic coating are important wire characteristics that should be taken into account when crossing a lesion. This article presents a few strategies for wire escalation as well as techniques to successfully cross and treat CTOs encountered when treating patients with CLI.
Typical wire lengths are 180 and 300 cm. A simple adage to keep in mind is: You can’t go wrong with long. This implies that using a 300-cm wire leaves the option to use any device off the shelf regardless of the disease location and access site. However, a shorter wire requires the lesion to be close to the access site or could require rapid exchange of equipment. For example, a right external iliac lesion can be effectively treated with a short wire (180 cm) with right common femoral artery access. On the other hand, a longer wire (300 cm) is needed to effectively treat a mid-left anterior tibial artery lesion from a contralateral retrograde right common femoral artery access.
The most common wire diameters are 0.014, 0.018, and 0.035 inch. Typically, smaller-diameter wires (0.014 and 0.018 inch) are used below the knee, and any diameter wire (0.014, 0.018, and 0.035 inch) can be used above the knee. Larger-diameter wires provide more support in crossing calcific CTOs and delivering treatment devices to their intended location (ie, 0.018 and 0.035 inch). It is important to keep in mind that larger-diameter wires are not benign, and dissections and perforations resulting from these wires are larger and more severe. Subintimal cannulation commonly occurs with a 0.035-inch wire and may warrant an alternative approach to cross the lesion (eg, retrograde) and/or the use of a reentry device. Smaller-diameter wires (0.014 and 0.018 inch) can be used just as effectively as larger wires for crossing and support. In fact, they can support larger-diameter devices and may be a safer alternative. Additionally, for extraluminal crossing, it is easier to pull a smaller-diameter wire back into true lumen and redirect without entering the false lumen again. Smaller-diameter wires are also better at finding and crossing microchannels than larger wires, which tend to bend and loop on themselves.
With respect to devices available for crossing and treatment, virtually all are able to go over a 0.014-inch guidewire. Careful attention is recommended when using an 0.014-inch wire looped on a 0.035-inch crossing catheter to avoid distal embolization. When the case necessitates escalating to a 0.018-inch wire, approximately one-third of devices can no longer be used. Similarly, escalating further to a 0.035-inch wire leaves few devices available for crossing and treatment. Compatibility of treatment devices with size of wire is an important consideration, and a wire should be carefully selected to fit the pathology of the lesions being treated.
Wires have a variety of tip strengths. Floppy-tipped wires (< 1 g) are used to cross nonocclusive blockages or traverse through a collateral to prevent trauma to the vessel. Stronger-tipped wires (ie, 3–40 g) are used to cross CTOs. Plaque morphology of the CTO helps determine which stronger-tipped wire to use. For example, if the plaque is heterogeneous, a 3- to 18-g wire may suffice. However, if the plaque is calcific, a 25- to 40-g wire may be warranted to traverse through the lesion. On occasion, if the interventionist knows the spatial location of the wire and is unable to cross a lesion, a 650-g tip (back end of an 0.018-inch wire) may be needed to successfully cross the occlusion. Additionally, the interventionist can increase the tip strength almost fourfold by bringing a support catheter within 10 mm from the tip of the wire. Wires with higher gram strength are not benign and should be used with extreme care.
Wires may have hydrophilic coatings to help penetrate microchannels or slide through tight lesions. Hydrophilic coatings may sacrifice tactile sensation to slide through lesions with ease. The coating on these wires may swell after prolonged time within the body and can lead to difficulty moving devices over the wire. Therefore, it is important to work quickly with these wires and/or switch them out when the lesions have been crossed. Noncoated wires are used in a rotating fashion to help drill through the caps of CTOs. Of note, noncoated wires are typically used with atherectomy devices instead of hydrophilic wires because hydrophilic wires tend to stick to the devices, and the forces generated by atherectomy devices tend to shear off the polymer coating.
Appropriate recognition of CTOs can influence guidewire selection. A systematic approach of pairing appropriate wires with techniques can optimize success and improve efficacy of treatments and minimize complications. A combination of CTO wires should be used in conjunction with a systematic approach to wire escalation. For example, a nonhydrophilic wire that is 300-cm long, has a 0.014-inch diameter, and an 18-g tip is used first. If unsuccessful, this can be escalated to a hydrophilic wire of the same size and gram strength. Next, a 30-g nonhydrophilic wire would be appropriate. Finally, the back end of an 0.018-inch nonhydrophilic wire (650-g tip) can be used.
There are several devices that complement wire crossing in CTOs. The first is the Spinr high-performance guidewire controller (Merit Medical Systems, Inc.), which cinches on the end of the wire and rotates the wire five times in a clockwise and then counterclockwise direction with a trigger mechanism. This movement decreases orthogonal friction, allowing the wire to traverse a long total occlusion with ease. Support catheters are important for wire support and can be used to increase the gram strength of the wire up to fourfold. Once the CTO wire crosses, the support catheter is used to inject dye and confirm true vessel entry. If it is not possible to cross the proximal cap of a CTO with a wire, a laser fiber can be used to cross the proximal cap of a CTO using the step-by-step technique. Once past the proximal cap, the wire is extended to cross the rest of the occlusion. The last device is extravascular ultrasound, which can be used to traverse CTOs by observing the outline of the vessel and manipulating the wire across the lesion (Figure 1).
There are some integral techniques that each interventionist should master to successfully cross CTOs. The first is the wrapping wire technique. Wires have a natural affinity for each other and tend to cross talk. This technique is especially useful in flush occlusions. Crossing the CTO from a retrograde location allows visualization of where the artery originates. A second wire can be taken from the antegrade access and wrapped around the retrograde wire to successfully cross from above.
The double-balloon technique, also known as the finger of god or CART technique, is used to successfully bring two wires into a common lumen when in two different subintimal planes. One balloon is traversed from a retrograde fashion, and a second balloon is traversed from an antegrade fashion in the CTO. The balloon tips approximate each other and are inflated simultaneously, which results in fenestration of the lumens such that the retrograde wire can enter the same lumen as the antegrade wire.4 The retrograde wire can then be advanced through the CTO. Once out of the CTO, the wire is externalized through an angled catheter in a flossing fashion. To do this, the interventionist must understand that a radiographic image is a two-dimensional image in a three-dimensional world. To successfully place the wire into the catheter without a snare, an angled catheter is pointed in the same direction as the angle on the wire. This serves as a guide to triangulate the position of the devices. The wire is then advanced into the catheter and externalized.
The controlled loop dissection technique is another useful strategy for crossing CTOs. A benign (< 3-g tipped, 0.014/0.018-inch wire) wire wraps around the tip of the catheter (0.018 or 0.035 inch). Figure 2 shows a fluoroscopic image of the controlled loop dissection technique crossing a CTO in an antegrade fashion. The controlled loop dissection technique uses a workhorse wire in an antegrade or retrograde fashion. The helical wrapping will maintain a relatively small profile of the wire/catheter combination, which will allow for controlled advancement of the wire/catheter combination, minimizing the size of the subintimal space. With the controlled loop dissection technique, the operator must be prepared to create subintimal recanalization of the CTO. The retrograde controlled loop dissection technique may be more forgiving because the dissection is retrograde in nature.
Another technique for crossing CTOs requires the use of the tip of the wire. Physicians use a heavy-gram-tipped wire to penetrate the CTO cap and navigate the course of the vessel. The operator may choose to use an angled or straight support catheter. A double-angled bend is created at the tip of the wire (Figure 3), which allows the wire point to penetrate the calcified, resistant CTO cap. The primary bend engages the CTO cap, while the secondary bend controls the direction of wire advancement. A triaxial system is used to support the wire, starting with a braided 5- to 7-F sheath and placing a 0.035-inch support catheter inside it, followed by placing a 0.014- or 0.018-inch lumen catheter within the 0.035-inch catheter. This triple layer supports a 0.014-inch CTO wire and directs the kinetic energy delivered from the physician’s hand to the wire tip. It is important to maintain a short distance (5–7 mm) between the CTO wire tip and the most inner supporting catheter of the triaxial system. For longer distances, the wire tends to buckle and wire support is lost.
There are many technical challenges in treating a patient with CLI. Successful revascularization can be limb saving for many patients and requires knowledge of patient anatomy, potential pathology, and proficiency with a variety of devices and techniques. CTOs and calcific lesions are often the most difficult to treat and require experience and a systematic approach that includes appropriate wire selection and escalation to maximize the possibility of procedural success.
Fadi Saab, MD
Interventional Cardiologist
Metro Health Hospital
Wyoming, Michigan
Clinical Assistant Professor of Medicine
Michigan State University CHM and COM
East Lansing, Michigan
Disclosures: None.
Vinayak Subramanian, BS
University of North Carolina School of Medicine
Chapel Hill, North Carolina
Disclosures: None.
The company is the world’s best Peripheral Guidewires supplier. We are your one-stop shop for all needs. Our staff are highly-specialized and will help you find the product you need.